The TM-108 Test Monitor monitors a vibration acceleration signal
and digitally displays the peak or rms acceleration and its corresponding
displacement. The TM-108 compares the acceleration and displacement
levels against the operating limits set and displayed on the front
panel push wheel switches.
Acceleration or displacement higher than the upper limits will
initiate a soft system shutdown via an optically isolated compressor
circuit. After the set minimum test level of acceleration has been
attained, an under-acceleration condition (loss of signal) will
also initiate a soft shut down. Shut down will also be automatically
initiated by a temporary or long term power line fault.
Immediate shut down can be further accomplished by the use of
the emergency stop button provided on the front panel or through
an external interlock fault loop (both normally closed and normally
open) which allows protection by external process or test faults.
|